Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Soltec TA200-938: (1 year) CAL VER /5520 DATE: 11-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 14 NUMBER OF LINES: 167 ============================================================================= # # Source: # Soltec TA100-438/TA200-938 Thermal Array Recorder Instruction Manual # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # Sub Soltec A11 (1 year) CAL VER /5520 # Sub Soltec A21 (1 year) CAL VER /5520 # Sub Soltec TJ1 (1 year) CAL VER /5520 # # Required Files: None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Connect UUT to AC power. 1.005 DISP 1.005 DISP Warm-up time: 30min - REQUIRED 1.005 DISP Temperature: 23 +/-2 degC 1.005 DISP Humidity: 55 +/-10% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 DISP Set ALL plug-ins LINE WIDTH knobs to OFF. 1.008 DISP This procedure will test the mainframe, and will then 1.008 DISP allow you to test the supported plug-in modules 1.008 DISP installed. (supported modules = A11, A21, & TJ1) 1.009 HEAD {FUNCTIONAL SELF-TEST} 1.010 DISP This recorder has a self-test built-in. 1.010 DISP 1.010 DISP For each test, set the TIME/DIV switch to proper range, 1.010 DISP turn the recorder OFF, hold the FEED button, and turn 1.010 DISP recorder ON. After 3 seconds release the FEED button. 1.011 SET TIME/DIVISION : 1 H 1.012 DISP Turn recorder OFF. 1.012 DISP Hold the FEED button. 1.012 DISP Turn recorder ON. 1.012 DISP Wait 3 seconds, the release the FEED button. 1.013 EVAL Is {the Software Firmware version printed (ST-1H)}? 2.001 SET TIME/DIVISION : 30 M 2.002 DISP Turn recorder OFF. 2.002 DISP Hold the FEED button. 2.002 DISP Turn recorder ON. 2.002 DISP Wait 3 seconds, the release the FEED button. 2.003 EVAL Do {all button LEDs light in sequence (ST-30M)}? 3.001 SET TIME/DIVISION : 10 M 3.002 DISP Turn recorder OFF. 3.002 DISP Hold the FEED button. 3.002 DISP Turn recorder ON. 3.002 DISP Wait 3 seconds, the release the FEED button. 3.003 EVAL Do {monitor channel LEDs light in various modes (ST-10M)}? 4.001 SET TIME/DIVISION : 5 M 4.002 DISP Turn recorder OFF. 4.002 DISP Hold the FEED button. 4.002 DISP Turn recorder ON. 4.002 DISP Wait 3 seconds, the release the FEED button. 4.003 EVAL Is {the complete character set printed (ST-5M)}? 5.001 SET TIME/DIVISION : 2 M 5.002 DISP Turn recorder OFF. 5.002 DISP Hold the FEED button. 5.002 DISP Turn recorder ON. 5.002 DISP Wait 3 seconds, the release the FEED button. 5.003 EVAL Are {button names printed when buttons are pressed (ST-2M)}? 6.001 SET TIME/DIVISION : 1 M 6.002 DISP Turn recorder OFF. 6.002 DISP Hold the FEED button. 6.002 DISP Turn recorder ON. 6.002 DISP Wait 3 seconds, the release the FEED button. 6.003 EVAL Are {print head lines at various intensities printed (ST-1M)}? 7.001 SET TIME/DIVISION : 30 S 7.002 DISP Turn recorder OFF. 7.002 DISP Hold the FEED button. 7.002 DISP Turn recorder ON. 7.002 DISP Wait 3 seconds, the release the FEED button. 7.003 EVAL Is {a triangle pattern printed (ST-30S)}? 8.001 SET TIME/DIVISION : 10 S 8.002 DISP Turn recorder OFF. 8.002 DISP Hold the FEED button. 8.002 DISP Turn recorder ON. 8.002 DISP Wait 3 seconds, the release the FEED button. 8.003 EVAL Is {all nomenclature printed when Time/Div rotated (ST-10S)}? 9.001 SET TIME/DIVISION : 5 S 9.002 DISP Turn recorder OFF. 9.002 DISP Hold the FEED button. 9.002 DISP Turn recorder ON. 9.002 DISP Wait 3 seconds, the release the FEED button. 9.003 EVAL Are {All side panel switch settings printed (ST-5S)}? 10.001 SET TIME/DIVISION : 2 S 10.002 DISP Turn recorder OFF. 10.002 DISP Hold the FEED button. 10.002 DISP Turn recorder ON. 10.002 DISP Wait 3 seconds, the release the FEED button. 10.003 EVAL Is {a solid black pattern printed (ST-2S)}? 11.001 SET TIME/DIVISION : 1 S 11.002 DISP Turn recorder OFF. 11.002 DISP Hold the FEED button. 11.002 DISP Turn recorder ON. 11.002 DISP Wait 3 seconds, the release the FEED button. 11.003 EVAL Does {the Module Monitor identify modules properly (ST-1S)}? 12.001 SET TIME/DIVISION : 100 uS 12.002 DISP Turn recorder OFF. 12.002 DISP Hold the FEED button. 12.002 DISP Turn recorder ON. 12.002 DISP Wait 3 seconds, the release the FEED button. 12.003 EVAL Is {current clock setting printed (ST-100uS)}? 13.001 OPBR Would you like to test the installed plug-ins? 13.002 JMPF 14.003 13.003 MATH M[1] = 1 13.004 LABEL DO_CHANNEL 13.005 MATH MEM = M[1] 13.006 HEAD CHANNEL [MEM] 13.007 RSLT =CHANNEL [MEM] 13.008 OPBR Would you like to test channel [MEM]? 13.009 JMPT 13.012 13.010 RSLT =Not tested 13.011 JMP 14.001 13.012 MEM2 Enter plug-in type (A11, A21, or TC1) 13.013 JMPL A11 ZCMPI(MEM2, "A11") 13.014 JMPL A21 ZCMPI(MEM2, "A21") 13.015 JMPL TC1 ZCMPI(MEM2, "TC1") 13.016 JMP 13.012 13.017 LABEL A11 13.018 RSLT =A11 - DC Voltage Module with Trigger (8-bit). 13.019 CALL Sub Soltec A11 (1 year) CAL VER /5520 13.020 JMP 14.001 13.021 LABEL A21 13.022 RSLT A21 - DC Voltage Module with Trigger (14-bit). 13.023 CALL Sub Soltec A21 (1 year) CAL VER /5520 13.024 JMP 14.001 13.025 LABEL TC1 13.026 RSLT TJ1 - Thermocouple Type-J. 13.027 CALL Sub Soltec TJ1 (1 year) CAL VER /5520 13.028 JMP 14.001 13.029 EVAL Dummy 14.001 MATH M[1] = M[1] + 1 14.002 JMPL DO_CHANNEL M[1] <= 9 14.003 END